Page 2 - Table of Contents
3 4 10 21 21 22 24 Table of Contents Definitions and Test Conditions System Performance Summaries Specifications and Characteristics Measurement Throughput Summary Options Test Set Specifications for Option 011 Accessories
Page 3 - Definitions and Test Conditions
3 Specifications describe the instrument’s warrantedperformance after a half-hour warm-up and overthe temperature range of 25° ± 5° C, unless other-wise stated. Specifications for frequencies above 3 GHz do not apply to instruments with Option 075(75-ohm impedance). Supplemental characteristics are ...
Page 4 - Configuration; Network analyzer; System dynamic range; GHz to 6 GHz; Measurement Port Characteristics; System Performance Summaries; Reflection uncertainty
4 Agilent 8753ES (50-ohm systems) 7-mm test ports Configuration Network analyzer 8753ES Standard and Option 006 Calibration kit 85031B Test-port cables 11857D System dynamic range 30 kHz to 50 kHz 70 dB 1 50 kHz to 300 kHz 90 dB 1 300 kHz to 16 MHz 100 dB 16 MHz to 3 GHz 110 dB Option 014 108 dB 3 G...
Page 5 - Transmission uncertainty
5 Agilent 8753ES (50-ohm systems) Type-N test ports Configuration Network analyzer 8753ESStandard and Option 006 Calibration kit 85032B Test-port cables 11857D System dynamic range 30 kHz to 50 kHz 70 dB 1 50 kHz to 300 kHz 90 dB 1 300 kHz to 16 MHz 100 dB 16 MHz to 3 GHz 110 dB Option 014 108 dB 3 ...
Page 7 - System Dynamic Range
Agilent 8753ES (75-Ohm Systems) Type-N test ports Configuration Network analyzer 8753ES Option 075 Calibration kit 85036B Test-port cables 11857B System Dynamic Range 30 kHz to 50 kHz 68 dB 1 50 kHz to 300 kHz 90 dB 1 300 kHz to 16 MHz 96 dB 16 MHz to 3 GHz 106 dB Measurement Port Characteristics Fr...
Page 10 - Agilent 8753ES Specifications
10 Test-Port Output Characteristics 7 Frequency characteristics Range 30 kHz to 3 GHz Option 006 30 kHz to 6 GHz Resolution 1 Hz Stability 9 ±7.5 ppm (0° to 55° C)±3 ppm/year Option 1D5 ±0.05 ppm (0° to 55° C),±0.5 ppm/year Accuracy ±10 ppm at 25° ±5° C Power range 2 -85 to +10 dBm Option 075/014 -8...
Page 11 - External source mode; Magnitude Characteristics
11 External source mode 1 (CW time sweep only) Frequency range 300 kHz to 6 GHz R-channel input requirements (typical) Power level 0 to -25 dBm Spectral purity Maximum <-30 dBc spurious input Residual FM <20 kHz Typical settling time 500 ms (automatic)50 ms (manual) Frequency readout accuracy ...
Page 14 - Agilent 8753ET Specifications
14 Test-Port Output Characteristics Frequency characteristics Range 300 kHz to 3 GHz Option 006 300 kHz to 6 GHz Resolution 1 Hz Stability 4 ±7.5 ppm (0° to 55° C)±3 ppm/year Option 1D5 ±0.05 ppm (0° to 55° C)±0.5 ppm/year Accuracy ±10 ppm at 25° ±5° C Power range -20 to +5 dBm-85 to +10 dBm (with O...
Page 15 - Magnitude characteristics; Display resolution; Phase Characteristics
15 External source mode 1 (CW time sweep only) Frequency range 300 kHz to 6 GHz R-channel input requirements 2 Power level 0 to -25 dBm Spectral purity Maximum spurious input <-30 dBcResidual FM <20 kHz Typical settling time 500 ms (automatic) 50 ms (manual) Frequency readout accuracy 2 0.1% (...
Page 16 - Measurement; Four display channels available.; Group Delay Characteristics; selectable; Source Control; Agilent 8753ET/ES Supplemental Characteristics
Measurement Number of display channels Four display channels available. Number of measurement channels Two primary and two auxiliary measurementchannels available. Measurement parameters 8753ET: Ref lection, transmission, A, B, R, A/R,B/R, A/B. Conversion to impedance or admittance.8753ES: S11, S21,...
Page 17 - The IF bandwidth is selectable from 6 kHz to 10 Hz.; Display Control
17 Sweep trigger Set to either continuous, single, group sweep, orexternal trigger. Set external trigger to take a com-plete sweep or to measure individual points in afrequency, power or list sweep. Power Set source power from -20 to +5 dBm for the 8753ET 1 or from -85 to +10 dBm for the 8753ES 2 . ...
Page 20 - This connector drives external VGA monitors.; General Characteristics
20 Front Panel Connectors 8753ES test ports (without Option 011) Connector type 7 mm, precision Impedance 50 ohms (nominal) 8753ES Option 011 test ports Connector type Type-N female Impedance 50 ohms (nominal) 8753ES Option 075 test ports Connector type Type-N female Impedance 75 ohms (nominal) 8753...
Page 21 - General Conditions; Operating conditions; Options; Environmental Characteristics
21 General Conditions RFI and EMI susceptibility: defined by VDE 0730,CISPR Publication 11, and FCC Class B Standards. ESD (electrostatic discharge): must be eliminatedby use of static-safe work procedures and an anti-static bench mat. The f lexible rubber keypad pro-tects key contacts from dust, bu...
Page 22 - Test Set Specifications for Option 011
22 Agilent 85046A/B S-Parameter Test Sets The 85046A/B S-parameter test sets provide thecapability to measure ref lection and transmissioncharacteristics (including S-parameters) of two-port devices in either direction with a single connection. The test sets are controlled from the8753ES Option 011 ...
Page 24 - Calibration Kits; mm short/open circuit; Accessories
24 Calibration Kits Vector accuracy enhancement procedures requirethat the systematic errors of the measurement sys-tem be characterized by measuring known devices(standards) on the system over the frequencyrange of interest. Return loss specifications or typical values are provided where available ...
Page 25 - Agilent 85038 Family of Calibration Kits; Agilent 11906 Family of Adapter Kits
25 Agilent 85038 Family of Calibration Kits The 85038A calibration kit contains open circuits,short circuits, and terminations with both maleand female 7-16 connectors. More economical ver-sions of this kit are available: the 85038M containsonly male devices, and the 85038F provides onlyfemale devic...
Page 27 - Spare 12 mil probes
27 Connectors 50-ohm type-N (f) to75-ohm type-N (m)standard Option 004, 50-ohm type-N (m) to 75-ohm type-N (f) Dimensions 14-mm D x 70-mm L(0.56 in x 2.75 in) Weight Net 0.1 kg (0.316 lb) 50-Ohm Accessory Kits The 11853A 50-ohm type-N and the 11854A 50-ohmBNC accessory kits provide the RF components...
Page 28 - Specifications
Agilent 8347A RF Amplifier The 8347A RF amplifier delivers increased power across a 300 kHz to 3 GHz frequency range. The 8347A provides leveled output powerwithout using an external coupler and detector,since these parts are built-in. The external ALCcan be directly connected to the external AM inp...