Agilent 8595E - Manual

Agilent 8595E

Agilent 8595E – Manual, read for free online in PDF format. We hope this helps you resolve any issues you may have. If you have further questions, please contact us through the contact form.

1 Page 1
2 Page 2
3 Page 3
4 Page 4
5 Page 5
6 Page 6
7 Page 7
Page: / 7

Table of Contents:

  • Page 3 – Amplitude Specifications; Amplitude Range; Average Continuous Power; Gain Compression; >10 MHz; Displayed Average Noise Level; kHz RBW; Nominal Dynamic Range
  • Page 5 – Audible Noise
  • Page 6 – Power Requirements; User Program Memory; Weight
  • Page 7 – Related Literature; General Purpose Information
Loading the manual

These specifications apply to the Agilent
Technologies 8591E, 8593E, 8594E, 8595E, and
8596E spectrum analyzers.

Specifications

All specifications apply over O°C to +55°C. The
analyzer will meet its specifications after 2 hours
of storage at a constant temperature, within the
operating temperature range, 30 minutes after the
analyzer is tumed on, and after CAL FREQ and
CAL AMPTD (and for the 8593E, 8595E, and
8596E CAL YTF) have been run.

Frequency Specifications

Frequency Range

8591E

50

9 kHz to 1.8 GHz

75

1 MHz to 1.8 GHz

8593E

9 kHz to 22 GHz

Option 026/027 9 kHz to 26.5 GHz

Band

LO harmonic = N

0

1

9 kHz to 2.9 GHz

1

1

2.75 GHz to 6.5 GHz

2

2

6.0 GHz to 12.8 GHz

3

3

12.4 GHz to 19.4 GHz

4

4

19.1 GHz to 22.0 GHz

4

4 (Opt. 026/027) 19.1 GHz to 26.5 GHz

8594E

dc coupled

9 kHz to 2.9 GHz

ac coupled

100 kHz to 2.9 GHz

8595E

dc coupled

9 kHz to 6.5 GHz

ac coupled

100 kHz to 6.5 GHz

8596E

dc coupled

9 kHz to 12.8 GHz

ac coupled

100 kHz to 12.8 GHz

Band

LO harmonic = N

0

1

9 kHz to 2.9 GHz (dc coupled)

0

1

100 kHz to 2.9 GHz (ac coupled)

1

1

2.75 GHz to 6.5 GHz

2

2

6.0 GHz to 12.8 GHz

Frequency Reference

(Opt. 004)

Aging

±2 x 10

-6

/year

±1 x 10

-7

/year

Temperature Stability

±5 x 10

-6

±1 x 10

-8

Initial Achievable Accuracy

±0.5 x 10

-6

±2.2 x 10

-8

Agilent 8590 E-Series
Portable Spectrum Analyzers

Data Sheet

"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.

Summary

Page 3 - Amplitude Specifications; Amplitude Range; Average Continuous Power; Gain Compression; >10 MHz; Displayed Average Noise Level; kHz RBW; Nominal Dynamic Range

3 Amplitude Specifications Amplitude specifications do not apply for Analog+mode and negative peak detector mode except asnoted in “Amplitude Characteristics.” Amplitude Range Displayed average noise levelto +30 dBm 8591E (Opt. 001) Displayed average noise level to+72 dBmV Maximum Safe Input Level (...

Page 5 - Audible Noise

5 Option Specifications Option 010 and 011 Tracking GeneratorFrequency Range 8591E 100 kHz to 1.8 GHz (Opt. 011, 75 Ω ) 1 MHz to 1.8 GHz 8593E, 94E, 95E, 96E 9 kHz to 2.9 GHz Output Level Range 8591E 0 to –70 dBm 8591E (Opt. 011) +42.8 to –27.2 dBmV 8593E, 94E, 95E, 96E –1 to –66 dBm Resolution 0.1 ...

Page 6 - Power Requirements; User Program Memory; Weight

6 Power Requirements ON (Line 1) 90 to 132 V rms, 47 to 440 Hz195 to 250 V rms, 47 to 66 HzPower consumption <500 VA; <180W Standby (Line 0) Power consumption <7 W User Program Memory 238 Kbytes non-volatile RAM Data Storage (nominal) Internal 24 traces or 32 states External 50 traces, 8 st...

Other Agilent Models

All Agilent Other