Contents 85032B/E iii 1. General Information Calibration Kit Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2 Kit Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ....
Page 5 - General Information
Page 6 - Calibration Kit Overview; Kit Contents; The 85032B calibration kit contains the following:
1- 2 85032B/E General Information Calibration Kit Overview Calibration Kit Overview The Agilent 85032B and 85032E type-N calibration kits are used to calibrate Agilent network analyzers up to 6 GHz for measurements of components with 50 Ω type-N connectors. Kit Contents The 85032B calibration kit co...
Page 7 - Adapters; Calibration Definitions; The calibration definitions can be:; Installation of the Calibration Definitions; The following options are available for the Agilent 85032B/E.; Equipment Required but Not Supplied; for ordering information
85032B/E 1-3 General Information Calibration Kit Overview Adapters Like the other devices in the kit, the adapters are built to very tight tolerances to provide good broadband performance. The adapters utilize a dual-beaded connector structure to ensure stable, repeatable connections. The beads are ...
Page 8 - Incoming Inspection
1- 4 85032B/E General Information Incoming Inspection Incoming Inspection Verify that the shipment is complete by referring to Figure 6-1 or Figure 6-2 . Check for damage. The foam-lined storage case provides protection during shipping. If the case or any device appears damaged, or if the shipment i...
Page 9 - Recording the Device Serial Numbers; for the 85032E. Recording the serial numbers will
85032B/E 1-5 General Information Recording the Device Serial Numbers Recording the Device Serial Numbers In addition to the kit serial number, the devices in this kit are individually serialized (serial numbers are labeled onto the body of each device). Record these serial numbers in Table 1-1 for t...
Page 10 - Clarifying the Sex of a Connector
1- 6 85032B/E General Information Clarifying the Sex of a Connector Clarifying the Sex of a Connector In this manual, the sex of calibration devices and adapters are referred to in terms of their connector interface. For example, a male open has a male connector. However, during a measurement calibr...
Page 11 - Specifications
Page 12 - Environmental Requirements; Temperature—What to Watch Out For; IMPORTANT
2-2 85032B/E Specifications Environmental Requirements Environmental Requirements Temperature—What to Watch Out For Changes in temperature can affect electrical characteristics. Therefore, the operating temperature is a critical factor in performance. During a measurement calibration, the temperatur...
Page 13 - Mechanical Characteristics; “Gaging Connectors” on page 3-6; Pin Depth; Connector Pin Depth
85032B/E 2-3 Specifications Mechanical Characteristics Mechanical Characteristics Mechanical characteristics such as center conductor protrusion and pin depth are not performance specifications. They are, however, important supplemental characteristics related to electrical performance. Agilent Tech...
Page 14 - NOTE; lists the typical; Pin Depth Limit
2-4 85032B/E Specifications Mechanical Characteristics NOTE The gages for measuring type-N connectors compensate for the designed offset of 5.26 mm (0.207 inch), therefore, protrusion and recession readings are in relation to a zero reference plane (as if the inner and outer conductor planes were in...
Page 15 - Electrical Specifications; Certification; “How Agilent Verifies the; Supplemental Electrical Characteristics; Supplemental Electrical Characteristics
85032B/E 2-5 Specifications Electrical Specifications Electrical Specifications The electrical specifications in Table 2-3 apply to the devices in your calibration kit when connected with an Agilent precision interface. Certification Agilent Technologies certifies that this product met its published...
Page 17 - Devices
85032B/E 3-1 3 Use, Maintenance, and Care of the Devices
Page 18 - Electrostatic Discharge; Figure 3-1 ESD Protection Setup
3-2 85032B/E Use, Maintenance, and Care of the Devices Electrostatic Discharge Electrostatic Discharge Protection against ESD (electrostatic discharge) is essential while connecting, inspecting, or cleaning connectors attached to a static-sensitive circuit (such as those found in test sets). Static ...
Page 19 - Visual Inspection; is recommended. However, not all defects that are; Look for Obvious Defects and Damage First; What Causes Connector Wear?; Inspect the Mating Plane Surfaces; . Look especially for deep scratches or
85032B/E 3-3 Use, Maintenance, and Care of the Devices Visual Inspection Visual Inspection Visual inspection and, if necessary, cleaning should be done every time a connection is made. Metal particles from the connector threads may fall into the connector when it is disconnected. One connection made...
Page 20 - Inspect Female Connectors; Cleaning Connectors; Use Compressed Air or Nitrogen; “Electrostatic Discharge”; Clean the Connector Threads
3-4 85032B/E Use, Maintenance, and Care of the Devices Cleaning Connectors Inspect Female Connectors Pay special attention to the contact fingers in the female center conductor. These can be bent or broken, and damage to them is not always easy to see. A connector with damaged contact fingers will n...
Page 21 - for part numbers for isopropyl alcohol and; Clean the Mating Plane Surfaces; conductor contact fingers by using short strokes.
85032B/E 3-5 Use, Maintenance, and Care of the Devices Cleaning Connectors Use a lint-free swab or cleaning cloth moistened with isopropyl alcohol to remove any dirt or stubborn contaminants on a connector that cannot be removed with compressed air or nitrogen. Refer to Table 6-3 on page 6-5 for par...
Page 22 - Gaging Connectors; Do not use the gages for precise pin depth measurements.; Connector Gage Accuracy; “Pin Depth” on page 2-3
3-6 85032B/E Use, Maintenance, and Care of the Devices Gaging Connectors Gaging Connectors The gages available from Agilent Technologies are intended for preventive maintenance and troubleshooting purposes only. (See Table 6-3 on page 6-5 for part number information.) They are effective in detecting...
Page 23 - When to Gage Connectors; Gage a connector at the following times:; Reading the Connector Gage
85032B/E 3-7 Use, Maintenance, and Care of the Devices Gaging Connectors When to Gage Connectors Gage a connector at the following times: • Prior to using a device for the first time: record the pin depth measurement so that it can be compared with future readings. (It will serve as a good troublesh...
Page 24 - Gaging Procedures; Gaging Male Type-N Connectors; “Visual
3-8 85032B/E Use, Maintenance, and Care of the Devices Gaging Connectors Gaging Procedures Gaging Male Type-N Connectors NOTE Always hold a connector gage by the gage barrel, below the dial indicator. This gives the best stability, and improves measurement accuracy. 1. Select the proper gage for you...
Page 26 - Gaging Female Type-N Connectors
3-10 85032B/E Use, Maintenance, and Care of the Devices Gaging Connectors Gaging Female Type-N Connectors NOTE Always hold a connector gage by the gage barrel, below the dial indicator. This gives the best stability, and improves measurement accuracy. 1. Select the proper gage for your connector. (R...
Page 28 - Connections; How to Make a Connection; Preliminary Connection; “Electrostatic Discharge” on page 3-2; CAUTION; Chapter 6; Torque Wrench Information
3-12 85032B/E Use, Maintenance, and Care of the Devices Connections Connections Good connections require a skilled operator. The most common cause of measurement error is bad connections. The following procedures illustrate how to make good connections. How to Make a Connection Preliminary Connectio...
Page 29 - Wrench Positions
85032B/E 3-13 Use, Maintenance, and Care of the Devices Connections Using a torque wrench guarantees that the connection is not too tight, preventing possible connector damage. It also guarantees that all connections are equally tight each time. 2. Prevent the rotation of anything other than the con...
Page 31 - How to Separate a Connection; Use an open-end wrench to prevent the device body from turning.; Handling and Storage; storage case when not in use.
85032B/E 3-15 Use, Maintenance, and Care of the Devices Handling and Storage How to Separate a Connection To avoid lateral (bending) force on the connector mating plane surfaces, always support the devices and connections. CAUTION Turn the connector nut, not the device body. Major damage to the cent...
Page 33 - Performance Verification
Page 34 - Introduction; Agilent verifies the specifications of these devices as follows:
4-2 85032B/E Performance Verification Introduction Introduction The performance of your calibration kit can only be verified by returning the kit to Agilent Technologies for recertification. The equipment required to verify the specifications of the devices in the kit has been specially manufactured...
Page 35 - Recertification; How Often to Recertify; Where to Send a Kit for Recertification; “Returning a Kit or
85032B/E 4-3 Performance Verification Recertification Recertification The following will be provided with a recertified kit: • a new calibration sticker affixed to the case • a certificate of calibration • a calibration report for each device in the kit listing measured values, specifications, and u...
Page 37 - Troubleshooting
Page 38 - Troubleshooting Process; Troubleshooting Flowchart
5-2 85032B/E Troubleshooting Troubleshooting Process Troubleshooting Process If you suspect a bad calibration, or if your network analyzer does not pass performance verification, follow the steps in Figure 5-1 . Figure 5-1 Troubleshooting Flowchart
Page 39 - Returning a Kit or Device to Agilent; Table 5-1 Contacting Agilent
85032B/E 5-3 Troubleshooting Returning a Kit or Device to Agilent Returning a Kit or Device to Agilent If your kit or device requires service, contact the Agilent Technologies office nearest you for information on where to send it. See Table 5-1 . Include a service tag (located near the end of this ...
Page 41 - Replaceable Parts
Page 42 - kit and; Replaceable Parts for the 85032B Calibration Kit
6-2 85032B/E Replaceable Parts Introduction Introduction Table 6-1 lists the replacement part numbers for items included in the 85032B calibration kit and Figure 6-1 illustrates each of these items. Table 6-2 lists the replacement part numbers for items included in the 85032E calibration kit and Fig...
Page 44 - Replaceable Parts for the 85032E Calibration Kit
6-4 85032B/E Replaceable Parts Introduction Figure 6-2 Replaceable Parts for the 85032E Calibration Kit Table 6-2 Replaceable Parts for the 85032E Calibration Kit Item No. Description Qty Per Kit Agilent Part Number Calibration Devices (50 Ω Type-N) 1 Male broadband load 1 00909-60009 2 Male combina...
Page 45 - Replaceable Parts—Items Not Included in the Calibration Kit
85032B/E 6-5 Replaceable Parts Introduction Table 6-3 Replaceable Parts—Items Not Included in the Calibration Kit Item No. Description Qty Agilent Part Number Connector Gages a (Type-N) a. To ensure you choose the correct gage, refer to, “Clarifying the Sex of a Connector” on page 1-6 . 1 Gage set (...
Page 47 - Standard Definitions
Page 48 - Standard Class Assignments; lists the classes of the devices in this calibration kit.; Standard Class Assignments
A-2 85032B/E Standard Definitions Standard Class Assignments Standard Class Assignments Class assignment organizes calibration standards into a format compatible with the error models used in the measurement calibration. A class or group of classes corresponds to the systematic errors to be removed ...
Page 49 - Blank Form; The standard class assignments listed in; Standard Class Assignments Blank Form
85032B/E A-3 Standard Definitions Standard Class Assignments Blank Form The standard class assignments listed in Table A-1 may be changed to meet your specific requirements. Table A-2 is provided to record the modified standard class assignments. Table A-2 Standard Class Assignments Blank Form Calib...
Page 50 - Nominal Standard Definitions; list typical calibration kit parameters used to specify the; Setting the System Impedance
A-4 85032B/E Standard Definitions Nominal Standard Definitions Nominal Standard Definitions Standard definitions provide the constants needed to mathematically model the electrical characteristics (delay, attenuation, and impedance) of each calibration standard. The nominal values of these constants...
Page 52 - The standard definitions listed in; Standard Definitions Blank Form
A-6 85032B/E Standard Definitions Nominal Standard Definitions Blank Form The standard definitions listed in Table A-3 may be changed to meet your specific requirements. Table A-4 is provided to record the modified standard definitions. Table A-4 Standard Definitions Blank Form System Z 0 a = ______...
Page 53 - Index
Index 85032B/E Index-1 A adapters , 1-3 part numbers , 6-5 Agilent Technologies, contacting , 5-3 alcohol isopropyl as cleaning solvent , 3-4 altitude , 2-2 B blank form standard class assignments , A-3 standard definitions , A-6 broadband loads , 1-2 C cal kit contents , 1-2 overview , 1-2 return t...