Page 3 - Contents
HW-G Instruction Manual Page 1 HW-G OZIM v1a Contents 1 Compliance ........................................................................................ 3 1.1 Compliance with FCC rules ........................................................... 3 2 Outline and Features ..............................
Page 5 - Compliance; Compliance with FCC rules
HW-G Instruction Manual Page 3 HW-G OZIM v1a 1 Compliance 1.1 Compliance with FCC rules Please note that this equipment generates, uses and can radiate radio frequencyenergy. This equipment has been tested and has been found to comply with thelimits of a Class A computing device pursuant to Sub-part...
Page 6 - Outline and Features
HW-G Instruction Manual Page 4 HW-G OZIM v1a 2 Outline and Features The HW-G series is a platform scale with a minimum resolution of 1/10000 . Type L scales have an LCD (Liquid Crystal Display) and use batteries as a powersource to provide portability. This type can also use an AC mains adaptor. Typ...
Page 7 - Unpacking
HW-G Instruction Manual Page 5 HW-G OZIM v1a 3 Unpacking Products HW-10KGL HW-10KGV Display Unit Base Unit HW-60KGL HW-60KGV Products Pan Display Uint HW-100KGL HW-200KGL HW-100KGV HW-200KGV Products Pan Caution Do not pull the load-cell cable. 5mm allen wrench All Accessories Refer to the accessrie...
Page 8 - Accessories and Options list; Accessories for HW-G series
HW-G Instruction Manual Page 6 HW-G OZIM v1a Type Products Accessories HW-10KGV Display unit coverInstruction manual Type V HW-60KGV Display unit cover HW-100KGV 5mm Allen wrench HW-200KGV Instruction manual HW-10KGL Display unit coverAC adaptorInstruction manual Type L HW-60KGL Display unit cover H...
Page 9 - Options List; Consumables
HW-G Instruction Manual Page 7 HW-G OZIM v1a Options List Order code or option name Accessories OP-03 RS232C interface / Relay output / Buzzer Connector JA:TCP0586 OP-04 RS422/485 interface / Relay output Connector TM:BLA9AC adaptor OP-06 Built-in printer for type V Paper PP156Ink ribbon ERC-05 AX-K...
Page 10 - Caution; Precautions for Installing the Scale; Precautions for Storing the Scale
HW-G Instruction Manual Page 8 HW-G OZIM v1a 4 Caution 4.1 Precautions for Installing the Scale Consider the following conditions to get the most from your scale.The best operation is where the temperature and relative humidity are stable, thescale is installed on a solid floor or bench and there is...
Page 11 - Installing the Scale
HW-G Instruction Manual Page 9 HW-G OZIM v1a 5 Installing the Scale This procedure includes all of the steps for installing the HW-G series scales Step 1 UnpackingRemove the weighing pan from the carton andplace to one side. Remove the base unit andcolumn from the carton, taking care not to pullon t...
Page 13 - Installing the Batteries for Type L
HW-G Instruction Manual Page 11 HW-G OZIM v1a 5.1 Installing the Batteries for Type L Step 1 If necessary turn off the displayand remove the AC adaptor. Step 2 Press and slide downwards to remove theexternal cover.Press the hook of the internal cover to theleft side and lift off the cover. Step 3 In...
Page 14 - Names
HW-G Instruction Manual Page 12 HW-G OZIM v1a 6 Names Base unit Display unit Angle adjustment Column Weigh pan CAL switch Calibrating the scale to weigh it correctly. Caution It is necessary to use classified mass. Battery Case Type L Bubble Spirit Level RS-232C Interface connector Leveling Foot
Page 15 - Display and Symbols
HW-G Instruction Manual Page 13 HW-G OZIM v1a Stability mark.When the current weight value is not changing, this markis displayed. The reading may now be taken. Zero point mark.W i t h n o t h i n g o n t h e p a n a n d p r e s s i n g t h e ZERO switch or with a mass on the pan and pressing the TA...
Page 19 - Basic Operation; Type V or Type L with AC adaptor
HW-G Instruction Manual Page 17 HW-G OZIM v1a 7 Basic Operation 7.1 Turning the Scale On/Off & Basic Weighing 7.1.1 Type V or Type L with AC adaptor Step 1 Ensure that the pan is empty. Step 2 Confirm that local voltage and receptacle type match your scale. Step 3 The scale turns on/off using th...
Page 20 - Type L with Batteries
HW-G Instruction Manual Page 18 HW-G OZIM v1a 7.1.2 Type L with Batteries Step 1 Install six new batteries. Refer to “5.1. Installing the batteries for Type L”. Step 2 Ensure that the pan is empty. Step 3 The scale turns on/off using the ON/OFF switch alternately. Step 4 Check the accuracy of weighi...
Page 23 - Counting Mode; Storing a Unit Mass
HW-G Instruction Manual Page 21 HW-G OZIM v1a Counting Mode 8 Counting Mode In counting mode the total weight of articles is converted to a count and assumesall articles have the same mass value. It is necessary to store a unit mass to count articles. 8.1 Storing a Unit Mass Step 1 Press the MODE sw...
Page 24 - Counting the number of articles
HW-G Instruction Manual Page 22 HW-G OZIM v1a Step 1 Press the MODE switch to display the unit PCS Step 2 Store the unit mass of the item. Refer to “8.1 Storing a Unit Mass” Step 3 Put the container only onto the pan. Press the TARE switch. Step 5 Put articles into the container and read the count. ...
Page 25 - Percentage Mode
HW-G Instruction Manual Page 23 HW-G OZIM v1a 9 Percentage Mode The percentage mode is to display a mass value in the unit of “%”. It is necessary to store a 100% mass value in advance to use this function. 9.1 Storing a 100% Mass Step 1 Press the MODE switch to display the unit %. Step 2 Press the ...
Page 26 - Reading Percentage
HW-G Instruction Manual Page 24 HW-G OZIM v1a 9.2 Reading Percentage Step 1 Press the MODE switch to display the unit %. Step 2 Store the unit mass of the article. Refer to “9.1 Storing a 100% Mass “ Step 3 If a container is needed, put the container only onto the pan and pressthe TARE switch. Step ...
Page 27 - 0 Accumulation Function
HW-G Instruction Manual Page 25 HW-G OZIM v1a This function counts the number of weighed items, calculates the total mass value andcan display the number and accumulated mass value. It is necessary to set the parameters of the “accumulation function ( f8 )” in the function table in advance to use th...
Page 30 - 1 Comparator Function; Comparator Sign
HW-G Instruction Manual Page 28 HW-G OZIM v1a 11 Comparator Function This function compares a display value with the upper limit (HI) and the lower limit (LO)and displays these results. It is necessary to set the “comparator function ( f6 0 ~ f6 7 )” parameters, upper limit value (HI) and lower limi...
Page 33 - Example 1; Case
HW-G Instruction Manual Page 31 HW-G OZIM v1a 11.2 Operation and Performance (Examples) Example 1 This example is set as follows:Function table f6 3 (If the current display value is not near-zero,the scale compares the display value with theupper limit value and the lower limit valueimmediately.) Up...
Page 34 - 2 Full/Dribble Batch Function
HW-G Instruction Manual Page 32 HW-G OZIM v1a 12 Full/Dribble Batch Function This function changes the scale to a filling machine which sub-divides a bulk product likegrain into loads of predetermined and virtually constant mass. It is necessary to set the parameter of the “comparator function ( f6 ...
Page 38 - 3 Simple Batch Function; Comparison Condition
HW-G Instruction Manual Page 36 HW-G OZIM v1a 13 Simple Batch Function This function compares a display value with the final value, preliminary value and zeroband for the full/dribble batch function. The result is indicated to zero band (LO indicator),full flow (HI indicator) and dribble flow (OK in...
Page 42 - The Gravity Acceleration Table
HW-G Instruction Manual Page 40 HW-G OZIM v1a 14.1 The Gravity Acceleration Table Amsterdam 9.813 m/s 2 Manila 9.784 m/s 2 Athens 9.800 m/s 2 Melbourne 9.800 m/s 2 Auckland NZ 9.799 m/s 2 Mexico 9.779 m/s 2 Bangkok 9.783 m/s 2 Milan 9.806 m/s 2 Birmingham 9.813 m/s 2 New York 9.802 m/s 2 Brussels 9....
Page 43 - The Complete Calibration Procedure
HW-G Instruction Manual Page 41 HW-G OZIM v1a 14.2 The Complete Calibration Procedure 14.2.1 Gravity Acceleration Correction Step1 Turn on the display.Open the rear cover of the display unit.Inside the display unit, press and hold the CAL switch to enter the calibration mode. Then Cal 0 is displayed...
Page 45 - 5 The Function Table; The Procedure for Setting Parameters
HW-G Instruction Manual Page 43 HW-G OZIM v1a 15 The Function Table The function table is provided to store and reference items that determine theperformance of the scale. Each item has a parameter.The parameters are maintained even without power applied. 15.1 The Procedure for Setting Parameters St...
Page 51 - Bit Format; Format 1
HW-G Instruction Manual Page 49 HW-G OZIM v1a 16.1 Data Format Bit Format Format 1 There are four headers for the type of data and weighing condition. Stable weighing data S T Unstable weighing data U S Stable counting data Q T Out of range (over) O L The weighing data consists of 9 characters inclu...
Page 52 - Format 2
HW-G Instruction Manual Page 50 HW-G OZIM v1a Format 2 There are four headers for the type of data and weighing condition. Stable weighing data S T Unstable weighing data U S Stable counting data Q T Out of range (over) O L The weighing data consists of 7 characters including decimal point and polar...
Page 54 - Tare; Cancel of tare
HW-G Instruction Manual Page 52 HW-G OZIM v1a 16.3 Command Mode The command mode is the function which can perform “output data”, “controlling the scale”and “setting parameters” by a command transmitted from a computer. Caution Allow at least 500 milliseconds between commands. 16.3.1 Command List Th...
Page 55 - Preset tare; Tare value is set and the net is displayed. The net mark is displayed.; Upper limit value; Lower limit value; Accumulation data output
HW-G Instruction Manual Page 53 HW-G OZIM v1a Preset tare Tare value is set and the net is displayed. The net mark is displayed. Template PT, [parameter] CommandResponse Net is displayed. Upper limit value An upper limit value is stored.Template HI, [parameter] CommandResponse The upper limit value ...
Page 56 - Final value; Preliminary value
HW-G Instruction Manual Page 54 HW-G OZIM v1a Final value The final value is stored. Template S0, [parameter] CommandResponse The value is stored. Preliminary value The preliminary value is stored.Template S1, [parameter] CommandResponse The value is stored. Zero band The zero band is stored.Templat...
Page 60 - Circuit
HW-G Instruction Manual Page 58 HW-G OZIM v1a Circuit Caution Please confirm that the AC adaptor type is correct for your local voltageand receptacle type. Selection switch for the RS-422/ RS-485 interface Selects either of RS-422 or RS-485 interface using a switch on the board. Installing OP-04 Ste...
Page 62 - Zero
HW-G Instruction Manual Page 60 HW-G OZIM v1a Zero The current mass value is set to the zero point, this is same as the ZERO switch. Template Z CommandReply The mass value becomes zero and zero point mark is displayed. Tare The current mass value becomes zero after placing a tare (receptacle, case, ...
Page 64 - Specification
HW-G Instruction Manual Page 62 HW-G OZIM v1a 17.3 Internal Printer for Type V (OP-06) It is necessary to set the parameter of the “Print mode ( f9 )” in the function table in advance. It is necessary to set the parameter of the “Date ( f16 )” on the function table in advance. Caution The date Param...
Page 65 - Installation of OP-06; Installation Method
HW-G Instruction Manual Page 63 HW-G OZIM v1a Installation of OP-06 Caution This option cannot be used with type L (Liquid Crystal Display) scales. Packed with the option OP-06 are the following items which are used to install theprinter into your scale. Reference No. Description Quantity 6 Printer ...
Page 70 - 9 Maintenance; Check Points Before Calling Maintenance
HW-G Instruction Manual Page 68 HW-G OZIM v1a 19 Maintenance Refer to “4. Caution” for use. Refer to “6.1. Display and Symbols “ and corresponding mode for displayed errorcode. Refer to “14. Calibration (Adjusting Scale)” for precision weighing. Periodically check the accuracy of weighing. Calibrate...